1. Adopting the principle of photoresistance method (shading method), it has the advantages of fast detection speed, strong anti-interference, high precision and good repeatability.
2. Adopting color high-definition LCD touch screen display, touch screen operation, and editable measurement program.
3. Built-in AS4059E 、AS4059F 、ISO4406-1987 、ISO4406-1999 、ISO4406-2017 、GJB420A-1996、GJB420B-2006 、GJB420B-2015 、NAS1638 、TOCT17216-71 、TOCT17216-01 、GB/T14039-02、NAS/ISO、SAE749D-1963、JTS-005、AETF1A and other particle contamination level standards, and can customize the channel according to user requirements.
4. Equipped with multiple detection modes including normal pressure and pressurized pressure, which can be flexibly selected according to sample viscosity; positive and negative pressure are adjustable to ensure the detection effect of complex samples.
5. Supports real-time display, storage (up to 100 groups) and traceable query of historical test data; it can print test results and histograms (under custom standards), and export data by connecting to a computer via RS232 interface, meeting the needs of data recording, tracing and analysis.
6. Built-in functions such as power-on cleaning, power-off cleaning, single cleaning and backflush blockage removal, which can specifically clean the sensor slit, syringe and pipeline; provides clear maintenance guidelines (such as printing paper replacement, air pump four-unit drainage/molecular sieve replacement) to reduce maintenance costs.
7. Equipped with a laser light-blocking sensor, a high-precision precision syringe (fragile parts are individually packaged with clear installation guidelines) and a magnetic stirrer (including magnetic isolation sheet to prevent ferromagnetic particle sedimentation); the positive and negative pressure adjustment of the air pump is stable to ensure the consistency of sampling and detection.
8. Supports three calibration methods including ACFTD, MTD and latex particle, with functions such as scale setting, noise measurement and moving window calibration (MOV); the instrument can be calibrated regularly to ensure long-term detection accuracy; the number of calibration points can be set between 4-16 to meet the needs of different detection ranges.
9. Convenient and Efficient Fault Diagnosis: The power-on self-test function can quickly identify abnormalities of core components such as sensors, samplers and valves, and provide clear fault handling guidelines; detailed troubleshooting and elimination steps are given for common faults such as sensor blockage to reduce downtime.
10. Suitable for clean laboratory environments (a clean bench is recommended); it has certain resistance to power supply interference; the working noise of the air pump can be reduced by padding foam at the bottom, meeting the operating conditions of conventional laboratories.